  data sheet zrf this ndir gas analyzer features a high accuracy, multiple functions and easy operation through use of a microprocessor. it also utilizes a mass flow detector noted for its high sensitivity and reliability. being housed in a 19 inch rack case suitable for mounting on a panel or a table top, this analyzer is applicable not only for measurement of environmental pollution but for use in various processes and/or experimental laboratories. outer dimensions (h x w x d): rack mounting type; 220 x 483 x 463mm panel flush mounting type; 220 x 443 x 463mm table top type; 232 x 443 x 463mm mass {weight}: approx. 20kg finish color: munsell 2.5y8.4/1.2 indication: 4 digit led for concentration 4 digit led for sub-indication output hold: output value before manual or automatic calibration is held. whether or not to effect hold function can be selected. sample gas condition: temperature; 0 to 50 c (due point of water vapor; less than 2 c) dust; less than 0.3 m pressure; less than 9.8kpa standard adjustment: calibration gas; dry n 2 balance interface compensation dew point of 2 c water vapor in n 2 warm up time: approx. 4 hours material of gas-contacting parts: sample cell; sus304, neoprene rubber infrared-ray transmitting window: caf 2 or sapphire internal tubing; teflon tube, silicone tube, toaron tube gas inlet/outlet, purge gas inlet size: rc1/4 (pt1/4 internal thread) or npt1/4 internal thread purge gas flow rate: 1 0.5 r /min it is necessary to purge the instrument interior when ambient air contains the corrosive gas etc. or the measuring range of co 2 is less than 0 to 50ppm. scope of delivery: analyzer, mounting bracket, test report, power fuse, cloth for cleaning infrared-ray transmitting window (1) use of a microprocessor provides high accuracy, multiple functions and easy operation. the built-in automatic calibrating function allows calibration of up to three components (option). the signal from a zirconia o 2 sensor (zfk3) or other o 2 meter enables output of an o 2 correction value (option). includes an alarm function providing an upper/ lower limit contact output (option). range can be changed over by external signal (option). zero and span calibration is accurate and easy by means of operating keys. ? self-diagnosis function is included. rs232c port available (option). (2) this analyzer utilizes a mass flow detector featuring high sensitivity and reliability. it is equipped with two standard ranges for a range ratio of up to 1:20. (3) addition of a zirconia o 2 sensor (zfk3) to the one/ two-component analyzer allows measurement of up to three components simultaneously. (4) besides the standard method of measurement, a sample switching system and differential flow system are also available. general items power supply: 100, 115 or 220v ac 10%, 50/60hz power consumption: 125va max. (220va max. when co/co 2 converter equipped) ambient temperature: ? to +45 c ambient humidity: 90%rh or less enclosure : steel casing, for indoor application microprocessor based ndir gas analyzer features specifications date EDS3-108E jun. 29, 2005
zrf 2 mounting method: mounted on 19 inch rack, or flush on panel, or on table top sample switching type this is an optimum analyzer for measurement of low concentrations or for eliminating the effects of interfering components. measuring system: non-dispersion infrared-ray absorption method, single light source ?double beam, sample switching system with in- tegrated zero air generator. measurable component: co (carbon monoxide) measuring range: output signal: 0 to 1v or 4 to 20ma dc, linear, step-like output which changes every 50 seconds repeatability: within 1% of full scale (within 2% of full scale)* linearity: within 1% of full scale zero drift: within 0.5% of full scale/week (within 1% of full scale/week)* span drift: within 1.5% of full scale/week (within 2.5% of full scale/week)* response time: within 120 seconds (for 90% response) measured gas flow rate: 2.0 0.1 r /min. note*: shows the value in case of 0 to 10ppm range. differential flow type this is an optimum gas analyzer for measurement in two modes, absolute concentration and concentration difference. measuring system: non-dispersion infrared-ray absorption method, single light source ?double beam, flow differential system with inte- grated zero air generator for co. measurable components and measuring range: remark: there are restrictions on the reference gas conditions. standard two-component analyzer measuring range: refer to table in page 11 and 12 output signal: 0 to 1v or 4 to 20ma dc (allowable load resistance 550 ? or less), linear repeatability: within 0.5% of full scale (within 1% of full scale)* linearity: within 1% of full scale zero drift: within 2% of full scale/week (within 3% of full scale/week)* span drift: within 2% of full scale/week (within 3% of full scale/week)* response time: 25 seconds max. (for 90% response )including gas substitution time; time differs with the length of sample cell measured gas flow rate: 0.5 0.25 r /min. (standard) note*: shows the value in case of 0 to 50ppm range. remark: 70% or more of the analyzer weight should be supported at the bottom of the case. (in case of mounting on panel or 19 inch rack, provide a support at the rear of casing). installation conditions: install the analyzer at a place not exposed to direct sunlight or the radiation from a high temperature object. avoid vibration, and select a clean place free of corrosive and/or combustible gases. if installing outdoor, provide a suitable casing or cover to protect the analyzer from wind, rain, etc. standard type measuring system: non-dispersion infrared-ray absorption method, single light source ?double beam measurable components and measuring range: standard single-component analyzer measurable component co co 2 no so 2 ch 4 cc r 4 n 2 o cf 3 chbrc r c 2 hf 3 c r -ochf 2 c 2 h 5 oh carbon monoxide carbon dioxide nitric oxide sulfur dioxide methane carbon tetrachloride nitrous oxide halothane ethlane ethyl alcohol 0 to 100 0 to 50 0 to 100 0 to 100 0 to 500 0 to 200 0 to 200 0 to 50 0 to 50 0 to 250 refer to table given in page 11 min. measuring range [ppm] measuring range measurable component (1st comp. + 2nd comp.) no+so 2 co+co 2 no+co co 2 +ch 4 nitric oxide + sulfur dioxide carbon monoxide + carbon dioxide nitric oxide + carbon monoxide carbon dioxide + methane 0 to 250/0 to 250 0 to 200/0 to 200 0 to 500/0 to 500 0 to 100/0 to 1000 refer to table in page 12 min. measuring range [ppm] measuring range 1st range [ppm] 0 to 10 0 to 25 0 to 50 0 to 100 2nd range [ppm] none, 0 to 20, 25, 50, 100 none, 0 to 50, 100 none, 0 to 100 none output signal: remark: linear output repeatability: within 0.5% of full scale linearity: within 1% of full scale zero drift: within 2% of full scale/week (wihtin 2% of full scale/day for 0 to 50ppm range) span drift: with in 2% of full scale/week (within 2% of full scale/day for 0 to 50ppm range) response time: 25 seconds max. (for 90% response) including gas substitution time consult to manufacturer } } consult to manufacturer } measurable component co 2 co carbon dioxide carbon monoxide ?50 to + 50 ?100 to + 100 0 to 100 0 to 200 0 to 250 0 to 500 0 to 1000 none, 0 to 200, 250, 500 none, 0 to 500 none, 0 to 500 1st range [ppm] 2nd range [ppm] co 2 co 1st range ?1 to + 1v dc 0 to 1v or 4 to 20ma dc 2nd range 0 to 1v dc 0 to 1v or 4 to 20ma dc
3 measured gas flow rate: 0.5 0.25 r /min. (reference gas) 0.5 0.25 r /min. (sample gas) optional specifications (these are added on request. refer to the "code symbols".) filter, flow checker: membrane filter and flow checker are built in. remark: the built-in membrane filter is a glass- fiber paper monitoring type. the pre- filter should be prepared separately. pump: a small two-throw electromagnetic pump is built in, so sample gas and reference gas can be sampled separately at the same time. co/co 2 converter (emission level calculation): this converter uses a special catalyst for converting efficiently into co 2 the co contained in sample gas which is used in the sample switching type etc. the converter is built in the analyzer. o 2 correction output (emission levels calculation): an exclusive o 2 sensor is used for cor- recting the measured gas concentration into the value at standard o 2 concentra- tion. for obtaining the nox and/or so 2 ex- haust standard value, zrf can measure the nox and/or so 2 concentration and simultaneously the residual oxygen con- centration in exhaust gas, and then cor- rect according to the following equation. (application of this equation is mandatory for the nox or so 2 exhaust standard.) c = 21 ?on ?cs 21 ?os where, c: concentration after o 2 cor- rection cs: nox or so 2 measured con- centration os: o 2 measured concentration on: o 2 standard concentration block diagram the o 2 measured signal is according to the exclusive o 2 sensor (type zfk) or external o 2 meter (0 to 1v dc/0 to 25% o 2 ). o 2 output signal: 0 to 1v or 4 to 20ma dc o 2 converted output signal: 0 to 1v or 4 to 20ma dc, linearity 2% of full scale; output can be provided for each of 1st and 2nd components alarm output: upper limit alarm; contact output 1c contact contact capacity 250v, 2a ac (resis- tive load) lower limit alarm; contact output 1c contact contact capacity 250v, 2a ac (resis- tive load) remote range changeover: range is changeable via external signal. range changeover input signal: 5v dc (minimum range selection at 5v input) range identification signal output: contact output 1a contact contact capacity 250v, 2a ac (resistive load) external output hold: output signal is held via external signal. output hold input signal: 5v dc average value output: average or moving average value output is available. average value is output every one or four hours. moving average value is output every one minutes it is averaged for one or four hours. (when select four hours average output the analyzer has only one average value output.) output signal; 0 to 1v or 4 to 20ma dc, linear automatic calibration: zero and span are automatically ca librated at the preset cycle. both of calibration gas and electromag- netic valve are not included. calibration channel: up to 3 components can be calibrated simultaneously. zero calibration point: fixed at 0% (zirconia o 2 meter allows setting zero points) span calibration point: 0 to 100% full scale calibration start: via built-in timer or remote start signal output hold at calibration: possible calibration gas flow mode: (1) zero gas (2) zero gas ?span gas 1 (3) zero gas ?span gas 1 ?span gas 2 (4) zero gas ?span gas 1 ?span gas 3 (o 2 ) (5) zero gas ?span gas 1 ?span gas 2 span gas 3 (o 2 ) calibration gas flow time: settable from 100 to 599 seconds calibration cycle: 1 to 99 hours (1-hour step) or 1 to 7 days (1-day step) calibration failure alarm: provided when fault occurs during auto calibration. contact output: during calibration; 1a contact, contact capacity 250v, 2a ac (resistive load) calibration failure; 1a contact, contact capacity 250v, 2a ac (resistive load) electromagnetic valve drive; 1a contact, contact capacity 250v, 2a ac (resis- tive load) remote start: remote start input signal; 5v dc square signal longer than 100msec. in duration communication interface: rs232c nox or so 2 signal o 2 signal 21 ?on ?cs 21 ?os o 2 corrected nox or so 2 [ppm]
zrf 4 in case of carbon monoxide measurement, a high perform- ance converter is provided to convert carbon monoxide into carbon dioxide. here, the carbon monoxide is eliminated and the gas is led into one cell of the high-sensitivity infrared analyzer. in the othe flow path, the gas is led directly into the other cell of the analyzer. the output of analyzer varies with the difference of the concentration of the carbon monoxide between two cells, eliminating the effects of interfering components. moreover, zero calibration can be held without zero stan- dard gas, flowing the same reference gas into both refe- rence and sample cell by activating changeover valve. the measured gas is divided into two, and in one of the flow paths, a high-performance converter is provided to convert carbon monoxide into carbonic acid gas. here, the carbon monoxide in the measured gas is eliminated and the gas is led into one cell of the high-sensitivity infrared analyzer. in the other flow path, the gas is led directly into the other cell of the analyzer. these flow paths are changed over via changeover valves sv1 and sv2 every 50 seconds by means of the changeover valve drive signal transmitted from the analyzer. by carrying out this changeover cyclically, the output of the analyzer varies with the concentration of the carbon monoxide in the measured gas. use of the obtained variation width as a measured value enables improving the s/n ratio and elimi- nating the effects of interfering components plus zero drift. sample gas outlet sample cell sample gas inlet tr i m mer reference cell interference compensating detector distribution cell (interference filter) infrared(single) light source motor rotary sector (one-point chopper) sector sync.signal o 2 meter (option) a/d cpu indicator d/a contact output ram rom rs232c external output hold remote range status range identification o 2 correction value av erage value alarm output 0 to 1v or 4 to 20ma dc operating keys indication gain changeover ac amplifier (option) (option) (option) pump zrf1,2 flow checker filter sample cell sample gas pump zrf4 flow checker filter reference cell sample cell reference gas sample gas sv co/co 2 conver ter pump zrf3 flow checker filter reference cell sample cell co sample gas sv 3 sv 1 sv 2 co/co 2 conver ter fundamental diagram description of differential flow system description of sample switching system optical sampling specification
5 code symbols y -- 2 y f r z- 21 14 15 16 17 18 19 20 9101 11213 2 1 345678 description measurable component so 2 co co 2 ch 4 no, (nox) other non-standard component power supply piping connection 100vac 50/60hz rc1/4 115vac 50/60hz npt1/4 220vac 50/60hz rc1/4 220vac 50/60hz npt1/4 1st component, 1st range 0 to 50ppm 0 to 100ppm 0 to 200ppm 0 to 250ppm 0 to 500ppm 0 to 1000ppm 0 to 2000ppm 0 to 5000ppm 0 to 1% 0 to 2% 0 to 5% 0 to 10% 0 to 20% 0 to 50% 0 to 100% other non-standard range measuring method single component analyzer structure/output signal table-top type 0 to 1v dc 19 inch rack mounting type 0 to 1v dc panel mounting type 0 to 1v dc table-top type 4 to 20ma dc 19 inch rack mounting type 4 to 20ma dc panel mounting type 4 to 20ma dc table-top type 0 to 1v dc rs232c 19 inch rack mounting type 0 to 1v dc rs232c panel mounting type 0 to 1v dc rs232c table-top type 4 to 20ma dc rs232c 19 inch rack mounting type 4 to 20ma dc rs232c panel mounting type 4 to 20ma dc rs232c optional components filter, flowchecker pump __ ? _ ?? optional function (1) auto calibration _ ? _ _ ? ? _ ? average value output _ _ ? _ ? _ ? ? remote range, alarm range identification, external hold _ _ _ ? _ ? ? ? a b d e p z 1 a b c d e f g h j k l m n p r z 0 2 3 4 a b c d e f g h j k l m y a b d f h k m 0 1 2 note(1): refer to table in page 11 for measuring range
zrf 6 (cont'd) y -- 2 y f r z- 21 14 15 16 17 18 19 20 9101 11213 2 1 345678 description optional function (2) o 2 indication/ correction note (2), (4) (external o 2 analyzer connection) _ ? _ ? _ note(2) any linearized 0 to 1v dc signal from external o 2 analyzer calibrated 0 to 25 vol% o 2 full scale is acceptable. note(3) standard measuring range of o 2 is 0 to 10% and 0 to 25%, double range. note(4) emission levels calculated with o 2 value output is available only when o 2 indication is selected. o 2 indication/ correction note (3), (4) (zfk connection) _ _ ? _ ? remote range, alarm range identification (for o 2 ) _ _ _ ? ? average value output time without 1-hour moving average value output 4-hours moving average value output other non-standard items 1st component, 2nd range without 2nd range 0 to 100ppm 0 to 200ppm 0 to 250ppm 0 to 500ppm 0 to 1000ppm 0 to 2000ppm 0 to 5000pppm 0 to 1% 0 to 2% 0 to 5% 0 to 10% 0 to 20% 0 to 50% 0 to 100% other non-standard range average value output 1st component instantaneous value o 2 correction __ ? _ _ ? o 2 standard value for emission leveles calculation (o 2 correction) none 4% 5% 6% 7% 10% 11% 12% 15% other non-standard items (speify within 0 to 19%) kind of measuring gas atmospheric gas flue gas converter exhaust gas other non-standard items non-standard spec. other non-standard items quick response note(8) quick response type is available only 0 to 20% range or more. note(5) refer to table in page 11 for measuring range 2nd range > 1st range range ratio : within 1 : 20 note(6) o 2 value must be informed for designation of "z". note(7) sample gas components must be informed for designation of "z". y a b c d y b c d e f g h j k l m n p r z 0 1 4 z y 4 5 6 7 a b c f z ey fy gy zz z a y a d
7 -- 2 f r z- 21 14 15 16 17 18 19 20 9101 11213 2 1 345678 description measuring method two component analyzer 1st component, 1st range 0 to 200ppm 0 to 250ppm 0 to 500ppm 0 to 1000ppm 0 to 2000ppm 0 to 5000ppm 0 to 1% 0 to 2% 0 to 5% 0 to 10% 0 to 20% 0 to 50% 0 to 100% other non-standard range measurable component no (nox) + so 2 co + co 2 no (nox) + co other non-standard components 2nd component, 1st range 0 to 200ppm 0 to 250ppm 0 to 500ppm 0 to 1000ppm 0 to 2000ppm 0 to 5000ppm 0 to 1% 0 to 2% 0 to 5% 0 to 10% 0 to 20% 0 to 50% 0 to 100% other non-standard range power supply piping connection 100vac 50/60hz rc1/4 115vac 50/60hz npt1/4 220vac 50/60hz rc1/4 220vac 50/60hz npt1/4 structure/output signal table-top type 0 to 1v dc 19 inch rack mounting type 0 to 1v dc panel mounting type 0 to 1v dc table-top type 4 to 20ma dc 19 inch rack mounting type 4 to 20ma dc panel mounting type 4 to 20ma dc table-top type 0 to 1v dc rs232c 19 inch rack mounting type 0 to 1v dc rs232c panel mounting type 0 to 1v dc rs232c table-top type 4 to 20ma dc rs232c 19 inch rack mounting type 4 to 20ma dc rs232c panel mounting type 4 to 20ma dc rs232c optional components filter, flowchecker pump __ ? _ ?? optional function (1) auto calibration _ ? _ _ ? ? _ ? average value output _ _ ? _ ? _ ? ? remote range, alarm range identification, external hold _ _ _ ? _ ? ? ? note(9) refer to table in page 12 for measuring range note(10) refer to table in page 12 for measuring range 2 f g h z c d e f g h j k l m n p r z 0 2 3 4 a b c d e f g h j k l m 0 1 2 y a c e g j l n c d e f g h j k l m n p r z
zrf 8 (cont'd) -- 2 f r z- 21 14 15 16 17 18 19 20 9101 11213 2 1 345678 description optional function (2) o 2 indication/ correction note (11), (13) (external o 2 analyzer connection) _ ? _ ? _ note(11) any linearized 0 to 1v dc signal from external o 2 analyzer calibrated 0 to 25 vol% o 2 full scale is acceptable. note(12) standard measuring range of o 2 is 0 to 10% and 0 to 25%, double range. note(13) emission levels calculated with o 2 value output is available only when o 2 indication is selected. non-standard spec. other non-standard items quick response o 2 indication/ correction note (12), (13) (zfk connection) _ _ ? _ ? remote range, alarm range identification (for o 2 ) _ _ _ ? ? 2nd component, 2nd range without 2nd range 0 to 500ppm 0 to 1000ppm 0 to 2000ppm 0 to 5000ppm 0 to 1% 0 to 2% 0 to 5% 0 to 10% 0 to 20% 0 to 50% 0 to 100% other non-standard range 1st component, 2nd range without 2nd range 0 to 500ppm 0 to 1000ppm 0 to 2000ppm 0 to 5000ppm 0 to 1% 0 to 2% 0 to 5% 0 to 10% 0 to 20% 0 to 50% 0 to 100% other non-standard range average value output time without 1-hour moving average value output 4-hours moving average value output other non-standard items average value output 1st component 2nd component instantaneous value o 2 correction instantaneous value o 2 correction without without without without ? ___ ? _ ? _ ? __ ? _ ? __ _ ?? _ _ ? _ ? __ ? _ __ _ ? o 2 standard value for emission leveles calculation none 4% 5% 6% 7% 10% 11% 12% 15% other non-standard items (speify within 0 to 19%) kind of measuring gas atmospheric gas flue gas converter exhaust gas other non-standard items note(14) 2nd range > 1st range range rate : within 1 : 20 note(15) 2nd range > 1st range range rate : within 1 : 20 note(17) sample gas components must be informed for designation of "z". note(18) quick response type is available only 0 to 20% range or more. note(16) o 2 value must be informed for designation of "z". y a b c d y e f g h j k l m n p r z y e f g h j k l m n p r z y a b c d e f g h y 4 5 6 7 a b c f z ey fy gy zz z a 0 1 4 z
9 -- 20 yy y y f r z- 21 14 15 16 17 18 19 20 910111 213 2 1 345678 description measuring method sample switching system 2nd range, 3rd range, 4th range 2nd range 3rd range 4th range without without without 20ppm without without 25ppm without without 50ppm without without 100ppm without without 25ppm 50ppm without 25ppm 100ppm without 50ppm 100ppm without 25ppm 50ppm 100ppm other non-standard range non-standard spec. other non-standard items power supply piping connection 100vac 50/60hz rc1/4 115vac 50/60hz npt1/4 220vac 50/60hz rc1/4 220vac 50/60hz npt1/4 1st range 0 to 10ppm 0 to 25ppm 0 to 50ppm 0 to 100ppm other non-standard items structure/output signal table-top type 0 to 1v dc 19 inch rack mounting type 0 to 1v dc panel mounting type 0 to 1v dc table-top type 4 to 20ma dc 19 inch rack mounting type 4 to 20ma dc panel mounting type 4 to 20ma dc table-top type 0 to 1v dc rs232c 19 inch rack mounting type 0 to 1v dc rs232c panel mounting type 0 to 1v dc rs232c table-top type 4 to 20ma dc rs232c 19 inch rack mounting type 4 to 20ma dc rs232c panel mounting type 4 to 20ma dc rs232c optional component pump co/co 2 converter _ ? ?? kind of measuring gas atmospheric gas other non-standard items optional function (1) auto calibration _ _ ? ? remote range, alarm range identification, external hold _ ? _ ? 3 v t a b z 0 2 3 4 a b c d e f g h j k l m 3 4 y y d y a y h y y 1 t a b 8 m n p z e y z z z b z measurable component co other non-standard components note(19) sample gas components must be informed for designation of "z".
zrf 10 -- 20 yy y y f r z- 21 14 15 16 17 18 19 20 910111 213 2 13 45678 description measuring method differential flow system 2nd range without 200ppm (available only for 1st range co 0 to 100ppm) 250ppm (available only for 1st range co 0 to 100ppm) 500ppm (not available for 1st range co 2 100ppm) 1000ppm (available only for 1st range co 2 100ppm) non-standard spec. other non-standard items optional components filter, flowchecker pump co/co 2 converter ___ ? __ ?? _ ? _ ? ??? power supply piping connection 100vac 50/60hz rc1/4 115vac 50/60hz npt1/4 220vac 50/60hz rc1/4 220vac 50/60hz npt1/4 structure/output signal table-top type 0 to 1v dc 19 inch rack mounting type 0 to 1v dc panel mounting type 0 to 1v dc table-top type 4 to 20ma dc 19 inch rack mounting type 4 to 20ma dc panel mounting type 4 to 20ma dc table-top type 0 to 1v dc rs232c 19 inch rack mounting type 0 to 1v dc rs232c panel mounting type 0 to 1v dc rs232c table-top type 4 to 20ma dc rs232c 19 inch rack mounting type 4 to 20ma dc rs232c panel mounting type 4 to 20ma dc rs232c note(20) when the 5th and 6th digit are "dq" or "ds", the output is available only voltage output. (1st range:?v to 1v, 2nd range:0 to 1v) kind of measuring gas atmospheric gas other non-standard items optional function (1) auto calibration _ _ ? ? remote range, alarm range identification, external hold _ ? _ ? 4 0 2 3 4 0 1 2 3 4 a b c d e f g h j k l m y y d y a y h y y c d e f e y z z z d q d s b b b c b d measurable component 1st range co 2 350 50ppm (carbon dioxide) 350 100ppm co 0 to 100ppm (carbon monoxide) 0 to 200ppm 0 to 250ppm note(22) sample gas components must be informed for designation of "z". note (21) available only for 5th digit "d" available only for 5th digit "b"
11 (standard single-component analyzer measuring ranges) remarks: (1) : standard measuring range (2) : consult with us regarding capability of manufacture, price and delivery period. (3) : outside of manufacturing range. ndir type infrared gas analyzer x 0 x 2 x 2.5 x 4 x 5 x 8 x 10 x 20 measurable component 2nd range x 0 x 2 x 2.5 x 4 x 5 x 8 x 10 x 20 x 0 x 2 x 2.5 x 4 x 5 x 8 x 10 x 20 x 0 x 2 x 2.5 x 4 x 5 x 8 x 10 x 20 x 0 x 2 x 2.5 x 4 x 5 x 8 x 10 x 20 50ppm 100ppm 200ppm 250ppm 500ppm 0.1% 0.2% 0.5% 1% 2% 5% 10% 20% 50% 100% 1st measuring range co co 2 no so 2 ch 4
zrf 12 (standard dual-component analyzer measuring ranges) remarks: (1) ? : available range for 1st measuring range. up to 1:20 possible for 2nd range. (max. 2000ppm for no analyzer) compbination of 1st, 2nd measurable components, measur- ing ranges: manufacture is possible as non-standard specifications even for some items not given in the table, so please consult to us and our distributor. 200ppm ???? ?? 250ppm ???? ??? 500ppm ?????? ?? 0.1% ??? ??? ??? 0.2% ???? ??? ??? 0.5% ? ??? ?? ??? 1% ??? ??? ? ??? 2% ??? ?????? 5% ??? ????? 10% ?? ????? 20% ? ????? 50% ????? 100% ????? 200 ppm 250 ppm 500 ppm 0.1 % 0.2 % 0.5 % 1 % 2 % 5 % 10 % 20 % 50 % 100 % co co 2 250ppm ?? 500ppm ?? no so 2 250 ppm 500 ppm garbage application co 2 : 0 to 50% ch 4 : 0 to 80% ( remarks: (1) ? : available range for 1st measuring range. (2) ? : available range for 2nd measuring range (max. range) for co and co 2 . 500ppm ?? 1000ppm ?? no co 500 ppm 1000 ppm
13 483 465 443 35 400 28 display panel operation keys po w er switch mounting bracket 220 146 support sample gas inlet sample gas outlet reference gas outlet* raference gas inlet* purge gas inlet external connection terminals (1st component) (2nd component) ( o 2 ) (auto cal) rs232c terminal power terminal 212 7 r3.5 435 213 +2 ? 436 +2 ? 70% or more of instrument mass (weight) should be supported at bottom of case. p anel cutout dimensions (in case of panel flush mounting) < rear view > outline diagram (unit:mm)
zrf 14 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 + - + - + + -- 0 to 1v or 4 to 20ma dc (instantaneous value) 0 to 1v or 4 to 20ma dc (moving average value) (option) output signal upper limit alarm contact output (option) lower limit alarm contact output (option) range identification contact output (option) f ault external output hold (option) remote range input (option) 1st component 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 + - + - + - + - + - o 2 output signal 0 to 1v or 4 to 20ma dc (instantaneous value) o 2 analyzer input signal 0 to 1v dc o 2 corrected value output (1st component) 0 to 1v or 4 to 20ma dc o 2 corrected value output (2nd component) 0 to 1v or 4 to 20ma dc range identification contact output (option) remote range input (option) upper limit alarm contact output (option) lower limit alarm contact output (option) o 2 (option) 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 + - auto calibration failure contact output remote start input a uto cal (option) 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 + - + - + - 0 to 1v or 4 to 20ma dc (instantaneous value) 0 to 1v or 4 to 20ma dc (moving average value) (option) output signal upper limit alarm contact output (option) lower limit alarm contact output (option) range identification contact output (option) remote range input (option) 2nd component under calibration contact output zero gas contact output span gas 1 contact output span gas 2 contact output span gas 3 contact output external connection diagram
15 for o 2 correction, the gas analyzer zrf can accept linealized 0 to 1v dc signal coming from analyzer calibrated 0 to 25% o 2 full scale. if the analyzer is not available, fuji can supply exclusive zirconia o 2 sensor model zfk3. measuring method: zirconia system measurable component and measuring range: repeatability: within 0.5% of full scale linearity: within 2% of full scale zero drift: within 1% of full scale/week span drift: within 2% of full scale/week response time: approx. 20 seconds (for 90% response) measured gas flow rate: 0.5 0.25 r /min remark: the zirconia system, due to its principle, may produce a measuring error due to relative concentration versus the com- bustible o 2 gas concentration. also, a corrosive gas (so 2 of 250 ppm or more, etc.) may affect the life of the sensor. gas inlet/outlet size: rc1/4 power supply: 90 to 126v ac or 200 to 240v ac, 50/60hz enclosure: steel casing, for indoor application indication: temperature indication (led) temperature alarm output: contact output 1a contact, contact capacity 220v, 1a ac (resistive load) outer dimensions (h x w x d): 140 x 170 x 190mm mass {weight}: approx. 3kg finish color: munsell 5y 7/1 measurable component 1st range 2nd range o2 oxygen 0 to 10vol% 0 to 25vol% exclusive zirconia o2 sensor (to be purchased separately) code symbols - 4 k f z1 y y 3 9 y 10 0 11 y 12 y 13 2 1 345678 description 3 y y 1 3 measuring method zirconia method power supply 90 to 126v ac 50/60hz 200 to 240v ac 50/60hz +- power fuse3a off 1234567 ac e out alm 800 7-m4 gas inlet 17 gas outlet 172.3 140 131 110 170 152 1234567 ac power 50/60hz supply e output to zrf alarm output +- 5 o xygen analyzer temp outline diagram (unit:mm) external connection diagram
zrf the product conforms to the requirements of the electromagnetic compatibility directive 89/336/eec as detailed within the technical construction file number tz734577. the applicable standards used to demonstrate compliance are : en 50081-1 : 1992 conducted and radiated emissions en 50082-1 : 1992 radiated immunity, esd and fbt zrf information in this catalog is subject to change without notice. printed in japan head office 6-17, sanbancho, chiyoda-ku, tokyo 102-0075, japan http://www.fesys.co.jp/eng sales div. international sales dept. no.1, fuji-machi, hino-city, tokyo, 191-8502 japan phone: 81-42-585-6201, 6202 fax: 81-42-585-6187 http: //www.fic-net.jp/eng caution on safety *before using this product, be sure to read its instruction manual in advance.


▲Up To Search▲   

 
Price & Availability of EDS3-108E

All Rights Reserved © IC-ON-LINE 2003 - 2022  

[Add Bookmark] [Contact Us] [Link exchange] [Privacy policy]
Mirror Sites :  [www.datasheet.hk]   [www.maxim4u.com]  [www.ic-on-line.cn] [www.ic-on-line.com] [www.ic-on-line.net] [www.alldatasheet.com.cn] [www.gdcy.com]  [www.gdcy.net]


 . . . . .
  We use cookies to deliver the best possible web experience and assist with our advertising efforts. By continuing to use this site, you consent to the use of cookies. For more information on cookies, please take a look at our Privacy Policy. X